LKG 610 - Electrical Safety Analyzer - 10 Lead ECG

Product Code : 610

LKG 610 Electrical Safety Analyzer is designed to test any piece of electrically operated patient care equipment. This full featured analyzer is microprocessor controlled and has a complete compliment of ten leads.

LKG 610 measures line voltage, outlet polarity, device current, earth leakage current, chassis resistance, and case leakage current. In addition, it performs lead ISO tests and point-to-point measurements.

All tests are quickly and easily performed by selecting the appropriate test function key. Measurement results are displayed on the large LCD screen with measurement units clearly indicated by an LED. Large switches change the polarity, open and close neutral, and lift the ground. Ten universal lead connectors accept both snaps and any size posts.

Calibrated test points are available to ensure perfomance accuracy and repeatability. Either AAMI or IEC 601-1 test loads may be selected. The receptacle is fused with an accessible 20A fuse and the tester itself is fused internally.

Delivered complete with a soft carry case and red test lead for resistance readings, LKG 610 is compact and light weight and provides the best combination of features, performance than any other comprable products in the market. 

 

Add to cart

    Fast, Accurate, and Portable
    Easy to Use
    Large LCD Display
    Accessible Line Fuse
    Calibrated Test Points
    Selectable Test Loads
    15 and 20 Amp Capabilities
    Point-to-Point Testing
    Uses any Test Cable
    10 Lead Universal Connector
    Complies with both AAMI and IEC 601-1 Test Loads
    LED Unit of Measure Indicator

     

    Display: 3 1/2 Digit LCD Display
    Measurement Range
    Resistance: 0 - 1999 milliOhms Leakage Current: 0 - 1999 microAmp Instrument Current: 0 - 19.99 Amps
    Accuracy
    Resistance: +/- 2% +/- 1 LSD Leakage: +/- 1% of FS DC to 1 KHz +/- 1 LSD +/- 2% of FS 1 KHz to 100 KHz +/- 4% of FS 100 kHz to 1 MHz Instrument Current: +/- 2% +/- 1 LSD
    Power Requirements
    110 VAC 50 - 60 Hz, 20 Amp or 220 VAC 10 Amp maximum rated for the test outlet and 0.1 Amps rated for the unit.
    Ambient Temperature
    Operating Range: 59 to 95 degrees F (15 to 35 degrees C) Storage Temperature: 0 to 122 degrees F ( -18 to 50 degrees C)
    Relative Humidity
    90% (max) at temperatures
    Calibrated Test Points
    Reference Ground: 1 Ohm Leakage: 200 microAmp

    General

    Size: 8.0" x4.5" x 2.0" (20.3 x 11.4 x 5.0 cm) 
    Weight:  2.5 lbs (1.1 kg)
    Controls: 12 Push Button Function Keys - Open / Closed, Neutral Switch Normal / Off / Reverse Polarity Switch, Momentary Ground Open Switch, 10 Universal Snap / Post ECG Connectors Test Load Selector Switch

     