Made in USA
ISO 9001:2015 Registered
SBA 8(a) Certified
Quotes
(0)

Defibrillator Analyzers

Back to Product Delta 3000

Products

Defibrillator Analyzers - Netech Model Delta 3000

Product Code : 3000

Delta 3000 Defibrilator Analyzer combines the complete testing functions of a Defibrillator Tester,  Transcutaneous Pacemaker Analyzer, Transvenous Pacemaker Analyzer and 12-Lead ECG / Arrhythmia Simulator into one compact, lightweight, and easy to-use instrument. No other single instrument matches the breadth of functions or delivers the performance testing power of the DELTA 3000.

Delta 3000 tests all Defibrillators including AEDs. It measures both monophasic and bi-phasic waveforms. It tests all external Pacemakers including transthoracic, transvenous, and AV sequential models. With 12 built-in user selectable test loads for transthoracic pacing testing, the built-in ECG / Arrhythmia Simulator allows performance testing of any ECG equipment.

Menu driven and easy to use, the Delta 3000 displays test measurements on the large four line LCD. A print key outputs the data to the RS-232 serial port.

Included with the Delta 3000 is a hard carrying case, various paddle adapters and interface cables. No other Defibrillator Analyzer can match the performance and value. Best of all Delta 3000 is FDA approved.

Add to cart

    Compact, Lightweight and Rugged
    Simple Menu Driven Operation
    Large 4-line LCD
    Simultaneous Display of All Test Measurements
    Dedicated Print Key for Immediate Documentaion

    RS-232  Interface

    Battery or AC Operation
    Two-Year Warranty

     

    Energy Measurement:

    Load: 50 Ohm +/- 1 %

    Range (High): 900 J Resolution: 1 Joule

    Range (Low): 0 - 100 Resolution: 0.1 Joules

    Accuracy: +/- 2% of the reading + 1 J for Hi Range or + 0.5 J for Low Range

    Maximum Voltage: 5200 V

    Maximum Current: 100 Amp (H)

    Sync. Time: 0 - 250 ms

    Waveform Output:

    Real Time: Output 1000 : 1 Stored.

    Discharge waveform expanded 200:1 and is repeated 8 sec. It can be viewed through the paddles and through the ECG High Level Output.

    Pacemaker Analyzer

    Transcutaneous:

    Test Load: 50 Ohms fixed; and variable 100 to 1000 Ohms (user selectable)

    Current: 1 to 200 mA Accuracy: +/- 5% + 0.5 mA

    Pulse Width: 0.5 to 80 ms Accuracy: +/- 1% + 0.1 ms

    Refractory: 20 - 500 ms Accuracy: +/- 2% or + 10 ms

    Atrial and Ventricular:

    Test Load: 500 Ohms

    Current: 1 to 25 mA Accuracy: +/- 5% + 0.5 mA

    Pulse Rate: 30 to 800 ppm. Accuracy: +/- 1% + 0.1 ms

    Refactory: +/- 20 - 500 ms

    Accuracy: +/- 2% or 10 ms

    Immunity Test: 50 / 60 Hz

    ECG Arrhythmia Simulator

    ECG: Normal Sinus Rhythm:

    Rate: 30, 60, 90, 120, 240

    Performance Waveforms:

    Sine, Square, Triangle and Pulse

    Rate: 0.5, 1, 2, 10, 15, 20, 25, 40 Hz

    Arrhythmias:

    VFBC, VFBF, VTAC, AFIB, PVC, PVC1, PVC2, BGY, RBB, AFIB, ATRFT, NSR

    Waveform Output:

    12 Lead ECG and Paddle and High Level Output Jacks

     

    General

    Power Requirements:

    Two 9V alkaline bBatteries or AC adapter

    Size:

    10" x 11" x 2.25" (25.4 x 27.9 x 5.7 cm)

    Weight:

    4 lbs (1.8 kg)

Download Brochure Download Manual
Part Number Description Price
650 Defibrillator; Transthoractic Pacer; Transvenous Analyzer Add to cart
652 Universal A/V Pacer Interface cable 2 Set Add to cart